Mobile Site Full Version
share:

Nonylphenol, Bisphenol A, 4-tert-Octylphenol, Nonylphenol Monoethoxylate, and Nonylphenol Diethoxylate on the Rxi®-5Sil MS by ASTM D7065-11 (Constant Flow)

PeakstR (min)Conc.
(µg/mL)
On-Column (ng)
1.Acenaphthene-D1013.6252.5
2.4-tert-Octylphenol15.4323.2
3.Nonylphenol15.9–17.016016
4.Phenanthrene-D1017.1252.5
5.4-Nonylphenol18.0323.2
6.Nonylphenol monoethoxylate18.5–20.032032
7.4-Nonylphenol monoethoxylate20.5323.2
8.Bisphenol A20.9323.2
9.Nonylphenol diethoxylate21.0–22.364064
10.Chrysene-D1223.4252.5
image
GC_EV1426
ColumnRxi®-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleMethod 525.2 internal standard mix (cat.# 31825)
AccuStandard® nonylphenol calibration standard (cat.# M-1626)
Diluent:Dichloromethane
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.:290 °C
Oven
Oven Temp.:50 °C (hold 2 min) to 320 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.44 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
13.045-5000.8
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9mm ID
Source Temp.:325 °C
Quad Temp.:150 °C
Solvent Delay Time:3.0 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesThe use of constant flow, rather than constant pressure, yields a faster analysis and more constant peak widths throughout the chromatogram.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.