Coal Tar PAHs on Rxi®-PAH (40 m x 0.18 mm x 0.07 µm)

PeakstR (sec)
1.Benzo[b]fluoranthene2,666.7
2.Benzo[k]fluoranthene2,673.2
3.Benzo[j]fluoranthene2,677.9
4.Benzo[a]fluoranthene2,698.9
image
GC_FF1235
ColumnRxi®-PAH, 40 m, 0.18 mm ID, 0.07 µm (cat.# 49316)
SampleNIST 1597a PAHs in a coal tar reference standard
Injection
Inj. Vol.:1 µL pulsed splitless (hold 1.47 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.1)
Inj. Temp.:275 °C
Pulse Pressure:80 psi (551.6kPa)
Pulse Time:1.5 min
Purge Flow:40 mL/min
Oven
Oven Temp.:80 °C (hold 1.5 min) to 340 °C at 4.9 °C/min (hold 10.44 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorFID @ 350 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Data Rate:10 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAHs

,

Rxi-PAH

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.