Cannabis Potency Compounds on Rxi®-5Sil MS

Peaks
1.Cannabidiol
2.Cannabichromene
3.delta-8-Tetrahydrocannabinol
Peaks
4.delta-9-Tetrahydrocannabinol
5.Cannabigerol
6.Cannabinol
image
GC_FF1253
ColumnRxi®-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620)
SampleCannabinoids standard (cat.# 34014)
Cannabichromene (cat.# 34092)
delta-8-Tetrahydrocannabinol (THC) (cat.# 34090)
Cannabigerol (cat.# 34091)
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium ® 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:150 °C (hold 0.1 min) to 330 °C at 25 °C/min (hold 0.7 min)
Carrier GasH2, constant flow
Flow Rate:1.6 mL/min
DetectorFID @ 350 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Medical marijuana

,

Cannabigerol

,

Cannabichromene

,

Cannabidiol

,

Cannabinol

,

THC

,

Medical cannabis

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.