Electronic Cigarette Liquid on Rtx®-VMS (Hydrogen/Fast Ramp)

PeakstR (min)
1.Methanol0.861
2.Ethanol0.905
3.Methylene chloride0.957
4.Propylene glycol1.433
5.Glycerin2.256
6.Nicotine3.030
image
GC_FF1257
ColumnRtx®-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample
Diluent:Methylene chloride
Conc.: Electronic cigarette liquid diluted 100:1
Injection
Inj. Vol.:1.0 µL split (split ratio 200:1)
Liner:Premium 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 260 °C at 54 °C/min (hold 0.15 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
Linear Velocity:67.2 cm/sec @ 100 °C
DetectorFID @ 250 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:H2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890A GC
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.