Mobile Site Full Version
share:

Cannabis Residual Solvents at 50 ppm Cutoff

Peaks
1.Ethyl mercaptan
2.Isobutane
3.Butane
4.Methanol
5.Pentane
6.Ethanol
7.Acetone
8.Isopropyl alcohol
9.n-Hexane
Peaks
10.Chloroform
11.Benzene
12.Heptane
13.Toluene
14.Ethylbenzene
15.m-Xylene
16.p-Xylene
17.o-Xylene
18.Dimethyl sulfoxide
image
GC_FS0516
ColumnRxi®-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
Sample
Diluent:Dimethyl sulfoxide
Conc.:10 ppm (concentration reflects 100 mg of sample at 50 ppm cutoff)
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1 mm straight (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:1.00 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Sample Temp.:140 °C
Platen temp equil. time:1.0 min
Sample Equil. Time:10.00 min
Vial Pressure:20 psi
Pressurize Time:5.0 min
Pressure Equilibration Time:0.20 min
Loop Pressure:15 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:35 °C (hold 1.50 min) to 175 °C at 40 °C/min to 225 °C at 30 °C/min
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 250 °C
Make-up Gas Flow Rate:45 mL/min
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

residual solvent

,

butane

,

solvents

,

624sil MS

,

cannabis

,

,

marijuana

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.