Mobile Site Full Version
share:

GC Multiresidue Pesticide Standard #9-OPP on Rxi®-5ms by GC-MS

PeakstR (min)
1.Mevinphos8.92
2.Phorate12.59
3.Fonofos13.73
4.Disulfoton14.05
PeakstR (min)
5.Methyl parathion15.06
6.Parathion (ethyl parathion)16.22
7.Triazophos19.59
8.Piperonyl butoxide20.50
image
GC_FS0608
ColumnRxi®-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleGC multiresidue pesticide standard #9-OPP (cat.# 32571)
Diluent:Toluene
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4.0 mm ID Precision® inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1555-5507
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Electron Energy:70 eV
Solvent Delay Time:5 min
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesReconstructed chromatogram from select ions.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.