Sulfur Compounds in Propane on Rt®-Silica BOND

Peaks
1.Carbonyl sulfide
2.Propane
3.Hydrogen sulfide
4.Carbon disulfide
5.Methyl mercaptan
image
GC_PC1273
ColumnRt®-Silica BOND, 30 m, 0.32 mm ID (cat.# 19785)
Sample
Conc.:6 ppm in 100% propane
Injectionsample valve
Sample Loop Vol.:250 µL
Inj. Temp.:250 °C
Oven
Oven Temp.:35 °C (hold 8 min) to 200 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorPFPD from OI Analytical @ 250 °C
Element Mode:sulfur
InstrumentThermo Trace GC
NotesThis valve/loop injection employed a split injection technique. Split flow was set to 40 mL/min.
AcknowledgementChromatogram courtesy of Jean-Louis Brix and Joeri Vercammen (Global Analyser Solutions, Belgium)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

sulfur

,

COS

,

H2S

,

CS2

,

methyl mercaptan

,

Silica BOND

,

,

mercaptan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.