Mobile Site Full Version
share:

Water Robustness Test on Rt®-Silica BOND

PeakstR (min)
1.Methane1.514
2.Ethane1.653
3.Ethylene1.698
4.Acetylene1.854
5.Propane2.032
6.Cyclopropane2.345
7.Propylene2.438
8.Propadiene2.594
PeakstR (min)
9.n-Butane3.151
10.1-Butene4.485
11.Methyl acetylene4.625
12.trans-2-Butene5.096
13.1,3-Butadiene5.267
14.Isobutylene5.267
15.cis-2-Butene5.544
16.iso-Pentane5.811
17.n-Pentane6.352
image
GC_PC1275
ColumnRt®-Silica BOND, 30 m, 0.32 mm ID (cat.# 19785)
SampleCustom hydrocarbon gas standard
Conc.: 1 mole percent in nitrogen
Injection
Inj. Vol.:15 µL split (split ratio 35:1)
Liner:Premium 2 mm ID straight inlet liner (cat.# 23313.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:120 °C (hold 25 min)
Carrier GasHe, constant flow
Linear Velocity:40 cm/sec
DetectorFID @ 260 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:10 Hz
InstrumentAgilent 7890A GC
NotesThe hydrocarbon mix was first analyzed to generate a baseline reference; then, ten separate 10 μL injections of water were made 1 minute apart using the same method. After the final water injection, the hydrocarbon mix was analyzed again to compare the chromatography before and after the water injections.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

selectivity

,

retenti

,

Silica BOND

,

,

Hydrocarbons

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.