Mobile Site Full Version
share:

Simulated Distillation on MXT-1HT SimDist by ASTM D7213

image
GC_PC1285
ColumnMXT-1HT SimDist, 5 m, 0.53 mm ID, 0.88 µm (cat.# 70131)
SampleRetention time standard for D7213 (C5-C72), Separation Systems, Inc.
Diluent:Carbon disulfide
Conc.:1 wt%
Injection
Inj. Vol.:1 µL PTV on column
Liner:Premium 1.7 mm ID on-column inlet liner (cat.# 23430.1)
Inlet Temp. Program:90 °C to 380 °C at 20 °C/min
Oven
Oven Temp.:40 °C to 380 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Flow Rate:12 mL/min
DetectorFID @ 390 °C
Make-up Gas Flow Rate:18 mL/min
Make-up Gas Type:N2
Hydrogen flow:43 mL/min
Air flow:400 mL/min
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

thick film column

,

SimDist

,

MXT

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.