Mobile Site Full Version
share:

Fatty Acids (C1-C7) on DiatoSorb-WHP Packed Column

Peaks
1.Acetic acid
2.Formic acid
3.Propionic acid
4.Isobutyric acid
5.n-Butyric acid
6.Isovaleric acid
7.n-Valeric acid
8.Isocaproic acid
9.n-Caproic acid
10.Heptanoic acid
image
GC_PC1289
Column10% FFAP, 1% H3PO4, 100/120 mesh on DiatoSorb-WHP, SilcoSmooth tubing, 2 m, 1/8 in OD, 2 mm ID
SampleFatty acids C1-C7
Diluent:Ether
Conc.:1,000 µg/mL
Injection
Inj. Vol.:8 µL on-column
Inj. Temp.:200 °C
Oven
Oven Temp.:125 °C (hold 2 min) to 175 °C at 10 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:30 mL/min
DetectorTCD @ 230 °C
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

packed columns

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.