Barbiturates on DiatoSorb WHP Packed Column

Peaks
1.Amobarbital
2.Pentobarbital
3.Secobarbital
4.Hexobarbital
5.Mephobarbital
6.Phenobarbital
image
GC_PC1292
Column3% OV-17 DA, 100/120 mesh on DiatoSorb WHP, SilcoSmooth tubing, 1 m, 1/8 in OD, 2 mm ID
SampleSecobarbital (cat.# 34038)
Hexobarbital (cat.# 34033)
Mephobarbital (cat.# 34034)
Phenobarbital (cat.# 34037)
Pentobarbital (cat.# 34036)
Conc.:0.2 mg/mL
Injection
Inj. Vol.:1 µL on-column
Liner:Packed column inlet adaptor kit for 1/4" injection ports (cat.# 21651)
Inj. Temp.:250 °C
Oven
Oven Temp.:210 °C (hold 12 min)
Carrier GasHe, constant flow
Flow Rate:20 mL/min
DetectorFID @ 275 °C
InstrumentAgilent 7890B GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Depressants

,

Packed Columns

,

Sedatives

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.