Mobile Site Full Version
share:

Biodiesel (B100) on MXT®-Biodiesel TG by ASTM D6584

image
GC_PC1293
ColumnMXT®-Biodiesel TG with 2 m Integra-Gap® (integrated retention gap), 16 m total length, 0.53 mm ID, 0.16 µm (cat.# 70289)
SampleBiodiesel (B100), derivatized
Diluent:Heptane
Injection
Inj. Vol.:1 µL cold on-column
Temp. Program:Oven track
Oven
Oven Temp.:50 °C (hold 1 min) to 180 °C at 15 °C/min to 230 °C at 7 °C/min to 380 °C at 30 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:3 mL/min
DetectorFID @ 380 °C
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Glycerides

,

Glycerin

,

Monoglycerides

,

Diglycerodes

,

Triglycerides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.