Mobile Site Full Version
share:

Permanent Gases on Capillary Rt-MSieve 5A

PeakstR (min)Conc. mol %
1.Hydrogen1.3584
2.Argon1.9763
3.Oxygen2.0282
4.Nitrogen3.5285
5.Methane4.7554
6.Carbon monoxide12.4995
image
GC_PC1331
ColumnRt-Msieve 5A, 30 m, 0.32 mm ID, 30 µm (cat.# 19722)
SamplePermanent gas standard
Diluent:Helium
Injection
Inj. Vol.:50 µL split (split ratio 20:1)
Liner:Topaz 2.0 mm ID straight inlet liner (cat.# 23313)
Inj. Temp.:150 °C
Oven
Oven Temp.:40 °C (hold 13 min)
Carrier GasHe, constant flow
Flow Rate:2.5 mL/min
DetectorTCD @ 200 °C
Make-up Gas Flow Rate:10 mL/min
Make-up Gas Type:He
InstrumentAgilent 7890B GC
NotesBaseline resolution between argon and oxygen is obtained using split injection. (R(Ar/O2) = 1.8)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

MolSieve

,

Natural Gas

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.