Topaz Liner Lot-to-Lot Reproducibility for QC Test Mix

PeakstR (min)Conc.
(µg/mL)
1.Acenaphthene (IS)7.5701.0
2.2,4-Dinitrophenol (DNP)7.6681.0
3.Carbaryl10.0200.5
4.Coumaphos13.7370.5
image
GC_QA0101
ColumnRxi-5ms, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13420)
SampleTopaz liner FID test mix
Diluent:Methylene chloride
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:Topaz 4.0 mm single taper liner with wool
Inj. Temp.:250 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:40 °C (hold 1 min) to 300 °C at 18.5 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:2 mL/min
DetectorFID @ 340 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:50 Hz
InstrumentAgilent 7890A GC
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

carbaryl

,

coumaphos

,

Topaz

,

liner

,

Topaz liner

,

wool

,

2

,

4-DNP

,

reproducibility

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.