Mobile Site Full Version
share:

Nonsteroidal Anti-Inflammatory Drugs (NSAIDS) on Raptor™ Biphenyl

PeakstR (min)Conc.
(µg/mL)
1.Piroxicam1.3925
2.Sulindac1.6025
3.Naproxen1.8525
4.Fenoprofen2.44100
5.Ibuprofen2.54200
6.Diclofenac2.9625
7.Indomethacin3.0825
image
LC_CF0561
ColumnRaptor™ Biphenyl (cat.# 9309A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:40 °C
Sample
Diluent:Water:acetonitrile (65:35)
Conc.:25-200 µg/mL
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.000.66535
4.000.64060
4.010.66535
5.000.66535
DetectorWater Acquity® PDA @ 270 nm
InstrumentWaters ACQUITY UPLC® H-Class
Download PDF