This chromatogram may contain analytes that are synonyms of your search terms.
Folate Deficiency Biomarkers in Human Plasma on Raptor HILIC-Si by LC-MS/MS
Peaks | tR (min) | Conc. (ng/mL) | Precursor Ion | Product Ion | |
---|---|---|---|---|---|
1. | 5-Formyl tetrahydrofolate | 2.23 | 50 | 474.2 | 327.0 |
2. | Folic acid![]() | 2.23 | 25 | 442.2 | 295.0 |
3. | 5-Methyltetrahydrofolic acid | 2.25 | 25 | 460.3 | 313.0 |

LC_CF0698
Column | Raptor HILIC-Si (cat.# 9310A5E) | ||||||||||||||||||
Dimensions: | 50 mm x 3.0 mm ID | ||||||||||||||||||
Particle Size: | 2.7 µm | ||||||||||||||||||
Temp.: | 30 °C | ||||||||||||||||||
Sample | |||||||||||||||||||
Diluent: | 20 mM Ammonium acetate in acetonitrile:water (80:20) containing 10 mg/mL 2-mercaptoethanol | ||||||||||||||||||
Inj. Vol.: | 5 µL | ||||||||||||||||||
Mobile Phase | |||||||||||||||||||
A: | 50:50 Water:acetonitrile, 20 mM ammonium acetate | ||||||||||||||||||
B: | 20:80 Water:acetonitrile, 20 mM ammonium acetate | ||||||||||||||||||
| |||||||||||||||||||
Max Pressure: | 344 bar | ||||||||||||||||||
Detector | MS/MS | ||||||||||||||||||
Ion Mode: | ESI+ | ||||||||||||||||||
Mode: | MRM | ||||||||||||||||||
Instrument | UHPLC | ||||||||||||||||||
Notes | Sample Preparation Method: 1. Aliquot 380 μL of human plasma (K2EDTA, 2x charcoal stripped) containing 100 μg/mL 2-mercaptoethanol and add 20 μL fortification solution. 2. Vortex for 2 min and centrifuge for 2 min at 4,000 rpm. 3. Condition EVOLUTE EXPRESS WAX 30 mg SPE plate (Biotage 604-0030-PX01) with 1 mL methanol and then equilibrate with 1 mL 2% formic acid in water. Apply vacuum to dry the plate completely. 4. Load 400 μL of sample onto the plate and apply vacuum to initiate the flow. 5. Wash the plate with 1 mL water. Apply vacuum to dry the plate completely. 6. Elute samples with 300 μL 5% ammonium hydroxide in methanol containing 10 mg/mL 2-mercaptoethanol. Apply vacuum for elution. 7. Evaporate extracts to dryness under nitrogen at 30 °C. 8. Reconstitute in 200 μL mobile phase B containing 10 mg/mL 2-mercaptoethanol. Note: The whole sample preparation process was performed under dim light. |
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.