Mobile Site Full Version
share:

This chromatogram may contain analytes that are synonyms of your search terms.

Explosives on Raptor ARC-18 by EPA Method 8330B

Peaks
1.HMX
2.RDX
3.1,3,5-Trinitrobenzene
4.1,4-Dinitrobenzene
5.1,3-Dinitrobenzene
6.3,5-Dinitroaniline
7.Nitrobenzene
8.Nitroglycerin
9.Tetryl
Peaks
10.2,4,6-Trinitrotoluene
11.2-Amino-4,6-dinitrotoluene
12.4-Amino-2,6-dinitrotoluene
13.2,4-Dinitrotoluene
14.2,6-Dinitrotoluene
15.2-Nitrotoluene
16.4-Nitrotoluene
17.3-Nitrotoluene
18.PETN
image
LC_EV0530
ColumnRaptor ARC-18 (cat.# 9314A1E)
Dimensions:100 mm x 3.0 mm ID
Particle Size:2.7 µm
Temp.:35 °C
Sample8330B Nitroaromatics and nitramine mix (cat.# 33204)
1,4-Dinitrobenzene (cat.# 33205)
Diluent:Water
Conc.:20 µg/mL
Inj. Vol.:3 µL
Mobile Phase
A:Water
B:Methanol
Time (min)Flow (mL/min)%A%B
0.001.07525
22.51.07525
DetectorUV/Vis @ 210/254 nm
Cell Temp:35 °C
Flow Cell Size:Semi-micro
InstrumentShimadzu UFLCXR
NotesThis is the recommended confirmation column for a two-column analysis. See LC_EV0531 for the primary column.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

8330B

,

explosives

,

ARC-18

,

,

Raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.