Mobile Site Full Version
share:

Phenylurea Herbicides on Raptor™ C18

PeakstR (min)
1.Tebuthiuron1.38
2.Thidiazuron1.72
3.Fluometuron2.77
4.Diuron3.09
5.Propanil4.21
6.Siduron4.39
7.Linuron4.84
8.Diflubenzuron6.42
image
LC_EV0533
ColumnRaptor™ C18 (cat.# 9304565)
Dimensions:150 mm x 4.6 mm ID
Particle Size:5 µm
Temp.:30 °C
SamplePhenylurea pesticide mixture (cat.# 32434)
Inj. Vol.:2 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.0027030
7.024555
7.0127030
10.027030
DetectorPDA @ 245 nm
InstrumentHPLC
NotesThis method is not EPA 532 tested or compliant.
Download PDF