Mobile Site Full Version
share:

Fat-Soluble Vitamins on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion
1.Retinol1.46269.493.3
2.Vitamin D23.52397.5379.6
3.Vitamin D33.58385.5367.5
4.γ-Tocopherol3.78417.6151.4
5.α-Tocopherol4.12431.7165.4
6.Vitamin K15.23451.4187.3
image
LC_FF0537
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:40 °C
Sample
Diluent:Methanol
Conc.:100 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid + 5 mM ammonium formate in water
B:0.1% Formic acid + 5 mM ammonium formate in methanol
Time (min)Flow (mL/min)%A%B
0.000.51090
4.00.50100
5.00.50100
5.010.51090
7.00.51090
Max Pressure:190 bar
DetectorABSCIEX API 4000
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Mode:MRM
InstrumentShimadzu UFLCXR
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

fat-soluble vitamine

,

vitamin D

,

vitamin E

,

vitamin A

,

vitamin K

,

vitamin

,

Raptor

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.