Mobile Site Full Version
share:

LC Multiresidue Pesticide Standard #4 on Raptor ARC-18 by LC-MS/MS

PeakstR (min)Precursor IonProduct Ion 1Product Ion 2
1.Cyromazine1.07167.085.0108.1
2.Pymetrozine1.35218.0105.079.0
3.Dinotefuran1.63203.0129.0157.0
4.Nitenpyram1.68271.1125.9224.9
5.Flonicamid1.89230.0203.1174.1
6.Imidacloprid2.19256.1175.1209.1
7.Acetamiprid2.40223.0126.056.1
8.Cymoxanil2.47199.0128.0111.0
9.Oxadixyl2.77279.0219.0132.0
10.Carbetamide2.79237.0118.0192.0
11.Secbumeton2.93226.2100.2170.2
12.Prometon2.93226.086.3184.3
13.Terbumeton2.97226.1114.1170.1
14.Imazalil2.97297.0159.069.0
15.Pyracarbolid3.04218.1125.197.1
16.Ametryn3.22228.168.1186.1
17.Pyrimethanil3.51200.0107.082.0
18.Spiroxamine isomer 13.56298.0144.0100.0
19.Metalaxyl3.56280.1220.1192.1
20.Spiroxamine isomer 23.62298.0144.0100.0
21.Chlorantraniliprole3.66483.9286.0453.0
22.Azoxystrobin3.92404.1372.0329.0
23.Furalaxyl3.96302.1270.1242.2
24.Fludioxonil3.97249.1229.1158.1
25.Halofenozide3.98331.1104.9275.0
26.Boscalid4.06342.9307.0139.9
27.Mandipropamid4.08412.3328.2356.2
28.Flutolanil4.10324.1262.165.0
29.Mepronil4.20270.191.0119.0
30.Myclobutanil4.25289.1125.170.2
31.Methoxyfenozide4.29369.1149.1313.2
32.Triadimefon4.33294.1197.269.3
33.Cyprodinil4.34226.093.0108.0
34.Mepanipyrim4.40224.1106.077.0
35.Fenhexamid4.46302.197.255.3
36.Fluoxastrobin4.46459.0427.0188.0
37.Butafenacil4.48492.0180.0331.0
38.Picoxystrobin4.78368.0145.1205.1
39.Rotenone4.83395.0213.1192.1
40.Tebufenozide4.86353.1133.0297.1
41.Dimoxystrobin4.87327.1116.1205.2
42.Carfentrazone ethyl4.89412.0346.0266.0
43.Kresoxim-methyl4.90314.1206.0116.0
44.Zoxamide5.06336.0187.1159.0
45.Famoxadone5.09392.2331.1238.0
46.Prochloraz5.14376.0308.070.1
47.Benalaxyl5.15326.1148.091.0
48.Clofentezine5.21303.0138.0102.0
49.Benzoximate5.28364.0199.1105.0
50.Trifloxystrobin5.50409.0186.0145.0
51.Fluazinam5.77464.8373.0338.1
52.Metaflumizone5.79507.0287.2267.1
53.Tebufenpyrad5.85334.0117.0145.0
54.Pyriproxifen5.91322.196.0227.1
55.Piperonyl butoxide5.93356.3176.9119.0
56.Quinoxyfen6.03308.0197.0161.9
57.Amitraz6.14294.0163.0122.0
58.Fenpyroximate6.36422.2138.1366.1
59.Abamectin B1b6.47876.6553.4291.0
60.Eprinomectin6.53914.6186.0154.0
61.Abamectin B1a6.61890.5305.2567.3
62.Fenazaquin6.68307.2161.057.2
63.Doramectin6.81916.6331.2593.4
64.Moxidectin6.81640.5498.3528.4
65.Ivermectin7.00892.6569.4551.4
Isomers are numbered sequentially based on elution order.
image
LC_FF0546
ColumnRaptor ARC-18 (cat.# 9314A12)
Dimensions:100 mm x 2.1 mm ID
Particle Size:2.7 µm
Temp.:50 °C
SampleLC multiresidue pesticide standard #4 (cat.# 31975)
Diluent:Water
Conc.:20 ng/mL
Inj. Vol.:5 µL
Mobile Phase
A:Water + 2 mM ammonium formate + 0.2% formic acid
B:Methanol + 2 mM ammonium formate + 0.2% formic acid
Time (min)Flow (mL/min)%A%B
0.000.4955
2.000.44060
4.000.42575
6.000.40100
7.000.40100
7.010.4955
9.500.4955
DetectorWaters Xevo TQ-S
Ion Source:Waters Zspray™ ESI
Ion Mode:ESI+
Mode:MRM
InstrumentWaters
NotesAutosampler Temp. = 5 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

multiresidue

,

pesticides

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.