Mobile Site Full Version
share:

Potency Analysis of a Commercially Available CBD Product on Raptor ARC-18 2.7 μm by LC-UV

PeakstR (min)
1.Cannabidiolic acid (CBDA)2.608
2.Cannabigerol (CBG)2.934
PeakstR (min)
3.Cannabidiol (CBD)3.101
4.Δ9-Tetrahydrocannabinol (Δ9-THC)5.851
5.Cannabichromene (CBC)7.315
image
LC_GN0581
ColumnRaptor ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:Raptor ARC-18 EXP guard column cartridge 2.7 µm (cat.# 9314A0250)
Temp.:30 °C
Sample
Diluent:25:75 Water:methanol
Conc.: 6.24 mg/mL (CBD)
Inj. Vol.:5 µL
Mobile Phase
A:Water, 5 mM ammonium formate, 0.1% formic acid
B:Acetonitrile, 0.1% formic acid
Time (min)Flow (mL/min)%A%B
0.001.52575
9.001.52575
DetectorUV/Vis @ 228 nm
InstrumentHPLC
Notes50 μL of a commerically available CBD product was diluted in 950 μL of 25:75 water:methanol. The sample was vortexed for 30 seconds at 3000 rpm. 400 μL of the sample was then filtered using a Thomson SINGLE StEP standard filter vial (PVDF, 0.2 μm, Restek cat.# 25895) prior to analysis. CBDA, CBG, CBD, Δ9-THC, and CBC were detected, but only CBD was quantified.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

cannabinoids

,

cannabis

,

potency

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.