Gas Purification

Restek Super Clean Gas Filter Kits

More Details
  • High-purity output ensures 99.9999% pure gas (at max. flow of 2 L/min).
  • Designed for easy, hassle-free cartridge changes.
  • Glass inside to prevent diffusion; polycarbonate housing outside for safety.

High-Capacity Oxygen Trap

More Details
  • Removes up to 135 ml of oxygen or 2 g of water.
  • Long lifetypically purifies more than 480 cu ft of inert gas.
  • Reduces oxygen to 15 ppb.
  • Maximum operating pressure: 250 psi (1,724 kPa).
  • Flow: 3 L/min @ 32 psi (221 kPa).

Capillary-Grade Hydrocarbon Trap

More Details
  • Packed with an extremely high surface area, baked coconut shell–based activated carbon.
  • Purged with ultra-high-purity helium; ready to use with any carrier gas including hydrogen.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • Maximum operating pressure: 250 psi (1,724 kPa).

Indicating Moisture Trap

More Details
  • Reduces water to less than 10 ppb; indicator changes from yellowish-green to blue at 5% relative humidity.
  • Prepurged for fast stabilization.
  • Reduces noise from high-sensitivity detectors.
  • Heavy-walled glass body prevents oxygen and water infusion.
  • 10 µm frit prevents microparticulate damage to needle valves and flow controllers.
  • Maximum operating pressure: 100 psi (689 kPa).

Indicating Oxygen Trap

More Details
  • Indicator changes from light green to grey as oxygen is trapped.
  • Heavy-walled glass body, protected by polycarbonate sleeve, prevents oxygen and water infusion.
  • Prepurged for fast stabilization.
  • 100 psi (689 kPa) maximum operating pressure.
  • Reduces oxygen to 0.1 ppm.
  • 10 µm frits at inlet and outlet.
  • Optimal flow rate: < 150 mL/min.

Replacement Chemical Traps and Parts

for Agilent GCs

More Details
  • Easy to install.
  • Attach to same fittings as original manufacturer's equipment.

Restek Click-On In-Line Super Clean Gas Traps and Connector Kits

More Details
  • High-purity output ensures 99.9999% pure gas.
  • Click-On fittings for easy, leak-tight cartridge changes; brass or stainless steel, 1/4" or 1/8".
  • Triple gas trap is ideal for purifying carrier gasit contains oxygen, moisture, and hydrocarbon scrubbers in one cartridge.
  • Fuel gas trap is ideal for purifying flame ionization detector (FID) fuel gases, removing both moisture and hydrocarbons.
  • Helium-specific triple gas trap is ideal for GC-MSit contains oxygen, moisture, and hydrocarbon scrubbers in one cartridge.

Click-On In-Line Super Clean Big Trap Replacement Filters

More Details
Note: Replacement filter onlyfittings sold separately.

Click-On In-Line Super Clean Big Trap Filter Kits

More Details
  • Largest-capacity traps availablerun longer without replacing.
  • "Quick connect" fittings with spring-loaded check valves allow filter replacement in seconds without tools or contamination.
  • Maximum operating pressure of 11 bar (160 psig).
  • High-purity output ensures 99.9999% pure gas (maximum flow of 8 L/min).
  • Triple gas trap is ideal for purifying carrier gas.
  • Helium-specific triple gas trap is made for GC-MSready to use within 15 minutes of installation.
  • Hydrocarbon trap removes hydrocarbon contamination from carrier gas to help eliminate ghost peaks.

Restek Super Clean Ultra-High Capacity Oxygen Filter

More Details
Restek Super Clean ultra-high capacity oxygen filter cartridges feature easy-to-read indicators.

The indicator code is shown on every trap so there is no confusion determining when a trap needs to be replaced.

Restek Super Clean Ultra-High Capacity Moisture Filter

More Details
Restek Super Clean ultra-high capacity moisture filter cartridges feature easy-to-read indicators.

The indicator code is shown on every trap so there is no confusion determining when a trap needs to be replaced.

Refillable Hydrocarbon Trap

More Details
  • Removes trace impurities from carrier gas.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • 60 µm frit prevents gas contamination by purifier particles.
  • Good for purge-and-trap systems.
  • Refillable and rechargeable.
  • Maximum operating pressure: 125 psig (862 kPa).
  • Maximum flow: 5 L/min.

High-Capacity Moisture Trap

More Details
  • Purged with ultra-high-purity helium; ready to use with any carrier gas including hydrogen.
  • Reduces water to less than 15 ppb.
  • Maximum operating pressure: 250 psi (1,724 kPa).
  • Maximum flow: 1.25 L/min.

Restek Gas Management System

More Details
  • Removes moisture, hydrocarbons, and oxygen from carrier gas, extending column lifetime.
  • Produces high-purity carrier gas for most applications.
  • Includes one each: moisture, hydrocarbon, and indicating oxygen trap.
  • Replacing traps is safe and easy.
  • Maximum flow: 150 cc/min for optimal performance.

Electronic Maintenance Indicators for Gas Filters

More Details
  • Audiovisual warning system ensures maintenance is done on time.
  • Progressively displays gas filter status, preventing unplanned downtime for maintenance.
  • Can be custom configured to your replacement/maintenance schedule.

Dynamic Duo (Restek Leak Detector and ProFLOW 6000 Flowmeter)

More Details
Protect your instrument and improve data quality with this powerful pair from Restek. Checking for leaks and verifying flows before you start running help you avoid costly problems later.

Click-On In-line Super Clean Ultra-High Capacity Oxygen Trap

More Details
Note: Replacement trap onlyfittings sold separately.

Click-On In-line Super Clean Ultra-High Capacity Moisture Trap

More Details
Note: Replacement trap onlyfittings sold separately.

Click-On In-line Super Clean Ultra-High Capacity Hydrocarbon Trap

More Details
Note: Replacement trap onlyfittings sold separately.

Restek ProFLOW 6000 Electronic Flowmeter

More Details
Flowmeters that can measure flammable gases are becoming mandatory due to the increased use of hydrogen in chromatography. With its Ex rating, the Restek ProFLOW 6000 meter is designed specifically with explosive and flammable gases in mind.

The Restek ProFLOW 6000 is the only flowmeter you need for any type of chromatography gas measurement because of its wide range of capabilities. The ProFLOW 6000 is an electronic meter capable of measuring volumetric flow for most gases. Real-time measurements can be made for various types of flow paths, including continually changing gas types. This portable unit is designed for easy handheld use, and the stand adds benchtop convenience.

Restek Electronic Leak Detector

New and improved! Prevent small leaks from causing big problems with a Restek leak detector.

More Details
  • Detects a broad range of gases and indicates leak severity with both an LED display and audible tone.
  • No more waiting for a full charge—can be operated during charging or used up to 12 hours between charges.
  • Charging kit includes both universal AC power adaptor and USB charging cable so you can charge anywhere, anytime.
  • Pinpoint very small gas leaks quickly and accurately before they cause damage and downtime.
  • Compact, handheld unit is easy to operate and convenient to use anywhere you need to check for leaks.

Click-On In-Line Super Clean Double Connector

More Details
Connects any Click-On trap to a Click-On indicator.

Click-On In-Line Super Clean Connectors

More Details
Click-On connectors allow you to change traps quickly, without introducing oxygen into your system.

Click-On In-Line Super Clean Indicator

More Details
  • Oxygen: green to grey
  • Moisture: beige to clear

Click-On In-Line Super Clean Replacement Gas Traps

More Details
Note: Replacement trap onlyfittings sold separately.

Helium-Specific Super Clean Filter and Kit

More Details
  • Specifically designed for purification of helium in GC-MS systems.
  • Traps are packed and conditioned using helium.
  • Uses standard single-position baseplate.

Restek Super Clean Gas Trapping System for LC-MS

More Details
A Super Clean quick-change cartridge system efficiently removes hydrocarbons from nitrogen!

  • Designed for easy, hassle-free cartridge changes.
  • Up to 20 L of hydrocarbon-free nitrogen per minute.
  • Filters connected in parallel to handle high flows for LC-MS.

Gas Pressure Gauge Kit

More Details
  • Use an in-line pressure gauge to indicate when a thermal gas purifier getter tube should be replaced.
  • Includes 1/8" tee and 0100 psi (0-689k Pa) gauge.

Restek Filter Baseplates

More Details
  • End fittings available in brass or stainless steel.
  • Baseplates fit all stand-alone Super Clean gas filters offered.

Hydrocarbon S-Trap

More Details
  • Removes hydrocarbons and other contaminants.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • Each individually activated to ensure maximum efficiency.
  • Fits in GC oven for easy thermal recharging.
  • Maximum operating pressure: 60 psi (414 kPa).

ECD Vent Trap

More Details
  • Reduces the release of hazardous materials from the ECD vent into the lab.
  • Includes connecting lines and mounting kit.

VICI Mat/Sen Gas-Specific Purifier Modules

More Details
  • Replace separate traps for removing impurities (e.g., moisture and hydrocarbons) with one multiple-bed purifier that is designed specifically for purifying helium, hydrogen, nitrogen, or air.
  • Reduce gas impurities from ppm to low ppb levels.
  • Decrease baseline noise and increase GC-MS sensitivity.
  • Pre-purged with the specified gas to shorten downtime.

High-Capacity Split Vent Trap

More Details
  • Reduces the release of hazardous materials from the capillary split vent into the lab.
  • Includes connecting lines and mounting kit.

Molecular Sieve S-Trap

More Details
  • Traps water vapor; increases column and oxygen trap lifetime.
  • Reduces baseline noise from sensitive detectors such as ECDs and mass spectrometers.
  • Activated and ready to use.
  • Reduces water to less than 1 ppm.
  • Maximum flow: 1 L/min.

Replacement Gas Filters for Restek Super Clean Gas Filter Kits

More Details
  • High-purity output ensures 99.9999% pure gas (at max. flow of 2 L/min).
  • Designed for easy, hassle-free cartridge changes.
  • Glass inside to prevent diffusion; polycarbonate housing outside for safety.
  • All traps measure 10 5/8" x 1 3/4" (27 x 4.4 cm).

Soft-Side Storage Case for Restek ProFLOW 6000 Electronic Flowmeter or Leak Detector

More Details
Optional soft-side storage case is ideal for storing your flowmeter or leak detector in smaller spaces, such as your tool box.

Wall Mounting Bracket

More Details
Baseplates can be mounted by using screws and the mounting holes on the baseplate, or by using this optional wall mounting bracket.

1/8-Inch to 1/4-Inch Tube End Reducer

More Details
To adapt 1/8" Super Clean baseplate fittings to 1/4" tubing, use 1/8" to 1/4" tube end reducers.

Replacement O-Rings for Cartridge Baseplates

More Details
Pack includes 10 large O-rings and 10 small O-rings.

Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us