share:

Sulfur Compounds on Rxi-1ms

Peaks
1.Hydrogen sulfide
2.Carbonyl sulfide
3.Methyl mercaptan
4.Ethyl mercaptan
5.Dimethyl sulfide
image
GC_AR00861
ColumnRxi-1ms, 30 m, 0.32 mm ID, 4.00 µm (cat.# 13396)
SampleHydrogen sulfide, carbonyl sulfide, methyl mercaptan, ethyl mercaptan, dimethyl sulfide
Diluent:Helium
Conc.:100 ppbv
Injectionsplitless
Inj. Temp.:30 °C
Oven
Oven Temp.:30 °C
Carrier GasHe, constant pressure
Linear Velocity:48 cm/sec @ 30 °C
DetectorSCD @ 800 °C
NotesInjection: sample loop (30°C), 1 mL splitless direct

Sample Storage and Tansfer:
SilcoCan air monitoring canister with Siltek-treated 1/4" valve (cat.# 24182-650); Sulfinert-treated gas sample loop, 1 cc (cat.# 22848); Sulfinert-treated gas sample loop, 10 cc (custom order)