share:

This chromatogram may contain analytes that are synonyms of your search terms.

TO-15 (HJ759) + PAMS on Rt-Alumina BOND/MAPD (by FID) and Rxi-624Sil MS (by MS) using Deans Switch

PeakstR (min)
1.Ethane7.68
2.Ethylene8.68
3.Propane10.36
4.Propylene15.58
5.Acetylene17.86
6.Dichlorodifluoromethane6.5
7.1,2-Dichlorotetrafluoroethane7.14
8.Isobutane7.19
9.Chloromethane7.35
10.trans-2-Butene7.84
11.n-Butane7.94
12.Vinyl chloride7.94
13.1,3-Butadiene8.16
14.cis-2-Butene8.35
15.1-Butene8.84
16.Bromomethane9.67
17.Chloroethane10.31
18.Isopentane10.77
19.Vinyl bromide11.26
20.Trichlorofluoromethane11.64
21.1-Pentene11.81
22.n-Pentane12.11
23.Ethanol12.68
24.trans-2-Pentene12.77
25.Isoprene13.21
26.cis-2-Pentene13.29
27.Acrolein13.65
28.1,1-Dichloroethene14.03
29.1,1,2-Trichlorotrifluoroethane14.17
30.Acetone14.27
31.2,2-Dimethylbutane14.27
32.Carbon disulfide14.83
33.Isopropyl alcohol14.99
34.Allyl chloride15.56
35.2,3-Dimethylbutane16.01
36.Methylene chloride16.05
37.2-Methylpentane16.14
38.Cyclopentane16.21
39.tert-Butyl alcohol16.49
40.Methyl tert-butyl ether16.94
41.trans-1,2-Dichloroethene16.95
42.3-Methylpentane16.97
43.1-Hexene17.56
44.n-Hexane17.81
45.1,1-Dichloroethane18.26
46.Vinyl acetate18.28
47.2,4-Dimethylpentane19.26
48.Methylcyclopentane19.46
49.2-Butanone19.69
50.cis-1,2-Dichloroethene19.73
51.Ethyl acetate19.78
52.Bromochloromethane20.28
53.Tetrahydrofuran20.3
54.Chloroform20.51
55.1,1,1-Trichloroethane20.88
56.2-Methylhexane20.89
57.Cyclohexane21.03
58.2,3-Dimethylpentane21.1
PeakstR (min)
59.Carbon tetrachloride21.21
60.3-Methylhexane21.28
61.Benzene21.63
62.1,2-Dichloroethane21.74
63.2,2,4-Trimethylpentane21.84
64.n-Heptane22.17
65.1,4-Difluorobenzene22.45
66.Trichloroethylene22.93
67.Methylcyclohexane23.38
68.1,2-Dichloropropane23.42
69.Methyl methacrylate23.43
70.1,4-Dioxane23.52
71.Bromodichloromethane23.89
72.2,3,4-Trimethylpentane24.08
73.2-Methylheptane24.37
74.3-Methylheptane24.64
75.cis-1,3-Dichloropropene24.65
76.Methyl isobutyl ketone24.85
77.Toluene25.25
78.n-Octane25.36
79.trans-1,3-Dichloropropene25.6
80.1,1,2-Trichloroethane25.95
81.Tetrachloroethene26.14
82.2-Hexanone26.23
83.Dibromochloromethane26.61
84.1,2-Dibromoethane26.84
85.Chlorobenzene-d527.5
86.Chlorobenzene27.55
87.Ethylbenzene27.64
88.n-Nonane27.74
89.m-Xylene27.81
90.p-Xylene27.81
91.o-Xylene28.38
92.Styrene28.4
93.Bromoform28.72
94.Isopropylbenzene28.86
95.4-Bromofluorobenzene29.14
96.1,1,2,2-Tetrachloroethane29.27
97.n-Propylbenzene29.42
98.1,2,3-Trimethylbenzene29.51
99.n-Decane29.55
100.p-Ethyltoluene29.57
101.2-Chlorotoluene29.6
102.1,3,5-Trimethylbenzene29.63
103.m-Ethyltoluene29.91
104.1,2,4-Trimethylbenzene30.14
105.1,3-Dichlorobenzene30.58
106.o-Ethyltoluene30.68
107.1,4-Dichlorobenzene30.70
108.Benzyl chloride30.82
109.m-Diethylbenzene30.86
110.p-Diethylbenzene30.99
111.n-Undecane31.02
112.1,2-Dichlorobenzene31.18
113.n-Dodecane32.33
114.1,2,4-Trichlorobenzene33.2
115.Hexachlorobutadiene33.32
116.Naphthalene33.62
image
GC_AR1172
ColumnRxi-624Sil MS, 60 m, 0.25 mm ID, 1.40 µm (cat.# 13869)
with zero dead volume Valco internal union (cat.# 20150)
SampleOzone precursor/PAMS mix (cat.# 26370)
75 Comp TO15 + NJ mix (cat.# 34392)
Conc.:10 ppbv
Injectionon-column
Oven
Oven Temp.:35 °C (hold 7 min) to 60 °C at 3 °C/min to 170 °C at 10 °C/min to 250 °C at 30 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:2.0 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
15.729-2263.7
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Source Temp.:230 °C
Quad Temp.:150 °C
Electron Energy:70 eV
Solvent Delay Time:5.7 min
Tune Type:BFB
Ionization Mode:EI
PreconcentratorMarkes Unity 2 + CIA
  Trap 1 Settings
Cooling temp.:5 °C
Desorb temp.:300 °C
Desorb flow:6 mL/min
Desorb time:180 sec
  Internal Standard
Purge flow:50 mL/min
Purge time:60 sec
Vol.:50 mL
ISTD flow:50 mL/min
  Standard
Size:400 mL
Purge flow:50 mL/min
Purge time:60 sec
Sample flow:100 mL/min
InstrumentAgilent 7890B GC & 5977A MSD
NotesDeans switch time: 6.489 min
Primary column to Deans switch: Rxi-624Sil MS 60 m x 0.25 mm x 1.40 μm (cat.# 13869), 2 mL/min He
Secondary column to FID: Rt-Alumina BOND/MAPD 30 m x 0.25 mm x 4 μm (cat.# 19781), 3 mL/min He
Transfer line to MS: intermediate-polarity deactivated 1 m x 0.1 mm (cat.# 10100), 3 mL/min He

FID Parameters
Temp.: 250 °C
H2 flow: 30 mL/min
Air flow: 400 mL/min
Makeup (N2) flow: 22 mL/min
Data rate: 5 Hz
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us