Acidic/neutral Drugs (Underivatized) on Rxi-5MS

Peaks
1.Methyprylon
2.Butalbital
3.Amobarbital
4.Meprobamate
5.Glutethimide
6.Phenobarbital
7.Methaqualone
8.Primidone
image
GC_CF00854
ColumnRxi-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleAcidic/neutral drugs
Diluent:methanol
Conc.:10 µg/mL each drug, 1 ng on column
Injection
Inj. Vol.:1.0 µL split (split ratio 10:1)
Liner:Gooseneck Splitless (4mm) Siltek
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 300 °C at 20 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:300 °C
Ionization Mode:EI
Scan Range:35-550 amu
NotesGooseneck Splitless (4mm) Siltek Liner obsolete, recommend catalog number 20798.