share:

This chromatogram may contain analytes that are synonyms of your search terms.

Blood Alcohol Analysis on Rtx-BAC Plus 2

PeakstR (min)Conc.
(µg/mL)
1.Acetaldehyde0.809100
2.Methanol0.861100
3.Ethanol1.034100
4.Acetone1.119100
5.Isopropanol1.181100
6.tert-Butanol1.30450
7.1-Propanol (n-Propanol)1.65850
image
GC_CF1154
ColumnRtx-BAC Plus 2, 30 m, 0.32 mm ID, 0.6 µm (cat.# 18006)
SampleBAC resolution control standard n-P (cat.# 36010)
BAC resolution control standard t-B (cat.# 36011)
Diluent:Water
Conc.: 50 µL of each standard were diluted in 900 µL water in a 20 mL headspace vial.
Injectionheadspace-loop split (split ratio 50:1)
Liner:1 mm ID straight inlet liner (cat.# 20972)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:3 min
Transfer Line Temp.:125 °C
Valve Oven Temp.:125 °C
Standby flow rate:50 mL/min
Sample Temp.:60 °C
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:1 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 3 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec @ 40 °C
DetectorFID @ 240 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesHeadspace concentrator courtesy of Teledyne Tekmar, Mason, OH.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

blood alcohol

,

headspace

,

BAC

,

alcohol

,

18000

,

18001

,

18002

,

18003

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us