share:

This chromatogram may contain analytes that are synonyms of your search terms.

BAC Resolution Control Standard t-B on Rtx-BAC Plus 1 and Plus 2 Column Set

PeaksConc.
(µg/mL)
BAC Plus 1 RT (min.)BAC Plus 2 RT (min.)
1.Methanol500.8010.741
2.Acetaldehyde500.8480.697
3.Ethanol501.0120.884
4.Isopropanol501.2381.006
5.Acetone501.3480.955
6.t-Butanol501.4681.108
image
GC_CF1161
ColumnsRtx-BAC Plus 1 30 m, 0.32 mm ID, 1.8 µm (cat.# 18004)
and Rtx-BAC Plus 2 30 m, 0.32 mm ID, 0.6 µm (cat.# 18006)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with Universal “Y” Press-Tight connector (cat.# 20405-261)
SampleBAC resolution control standard t-B (cat.# 36011)
Conc.: 50 µL of standard was diluted with 950 µL water in a 20 mL headspace vial.
Injectionheadspace-loop split (split ratio 50:1)
Liner:1 mm straight inlet liner (cat.# 20972)
Headspace-Loop
Inj. Port Temp.:200 °C
Instrument:Tekmar HT3
Inj. Time:1 min
Transfer Line Temp.:125 °C
Valve Oven Temp.:125 °C
Sample Temp.:60 °C
Sample Equil. Time:5 min
Vial Pressure:30 psi
Pressurize Time:2 min
Loop Pressure:20 psi
Loop Fill Time:1 min
Oven
Oven Temp.:40 °C (hold 3 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 240 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
InstrumentAgilent/HP6890 GC
NotesThe Rtx-BAC Plus 1 and Plus 2 columns were connected to the injection port using a ~12 inch section of guard column between the injection port and the Universal Y Press-Tight connector.

Headspace concentrator courtesy of Teledyne Tekmar, Mason, OH.
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us