share:

Hexamethylenediamine on Stabilwax-DB

Peaks
1.Cyclohexane
2.Hexamethyleneimine
3.1,4-Diaminobutane
4.Pentamethylenediamine
5.1,2-Diaminocyclohexane
6.1,5-Diamino-2-methylpentane
Peaks
7.Aminomethylcyclopentylamine
8.Hexamethylenediamine
9.6-Aminocapronitrile
10.n-Valeramide
11.Adiponitrile
12.bis-Hexamethylenetriamine
image
GC_CH00302
ColumnStabilwax-DB, 30 m, 0.32 mm ID, 0.25 µm (cat.# 10824)
SampleNeat hexamethylenediamine (HMD)
Conc.: 10 to 1,000 ng/component on-column
Injection
Inj. Vol.:0.4 µL Direct
Inj. Temp.:250 °C
Oven
Oven Temp.:95 °C (hold 6 min) to 235 °C at 7 °C/min (hold 4 min)
Carrier GasH2, constant pressure
Linear Velocity:40 cm/sec
DetectorFID @ 250 °C
NotesFID sensitivity: 2 x 10-11 AFS
Download PDF