Chlorofluorocarbons on Rtx-200
Peaks | |
---|---|
1. | Freon® 21 |
2. | Freon® 11 |
3. | Freon® 123 |
4. | Freon® 123a |
5. | Freon® 113 |
6. | Dichloromethane![]() |

GC_CH00322
Column | Rtx-200, 105 m, 0.53 mm ID, 3.00 µm (cat.# 15091) |
Sample | Mix of Freon compounds |
Conc.: | 500 µg/mL |
Injection | |
Inj. Vol.: | 1.0 µL splitless (hold 0.5 min) |
Inj. Temp.: | 220 °C |
Oven | |
Oven Temp.: | 35 °C |
Carrier Gas | He, constant flow |
Linear Velocity: | 20 cm/sec @ 35 °C |
Detector | MS |
Mode: | Scan |
Scan Rate: | 1.5 scans/sec |
Notes | MSD: TIC mode |