share:

This chromatogram may contain analytes that are synonyms of your search terms.

Styrene Impurities Rtx-1701

Peaks
1.1,3-Butadiene
2.Butene
3.Acrylonitrile
4.Diethylhydroxylamine
5.Toluene
6.Vinylcyclohexene
7.Ethylbenzene
8.m-Xylene
9.o-Xylene
10.Styrene
11.Cumene
image
GC_CH00357
ColumnRtx-1701, 30 m, 0.53 mm ID, 3.00 µm (cat.# 12085)
Sample95% pure styrene
Injection
Inj. Vol.:500 µL split
Inj. Temp.:150 °C
Split Vent Flow Rate:40 mL/min
Oven
Oven Temp.:40 °C (hold 10 min) to 150 °C at 12 °C/min (hold 15 min)
Carrier GasHe, constant pressure
Linear Velocity:20 cm/sec @ 40 °C
DetectorFID @ 150 °C
NotesFID sensitivity: 16 x 10-11 AFS
AcknowledgementPermission to publish this chromatogram granted by Copolymer Rubber and Chemical Corp.