share:

This chromatogram may contain analytes that are synonyms of your search terms.

Volatile Organics by EPA Method 502.2 on Rtx-502.2

Peaks
1.Dichlorodifluoromethane
2.Chloromethane
3.Vinyl chloride
4.Bromomethane
5.Chloroethane
6.Trichlorofluoromethane
7.Freon 113
8.1,1-Dichloroethene
9.Allyl chloride
10.Methylene chloride
11.Methyl tert-butyl ether
12.trans-1,2-Dichloroethene
13.1,1-Dichloroethane
14.Chloropropene (40 ppb)
15.2,2-Dichloropropane
16.cis-1,2-Dichloroethene
17.Chloroform
18.Bromochloromethane
19.1,1,1-Trichloroethane
20.1,1-Dichloropropene
21.Carbon tetrachloride
22.1,2-Dichloroethane
23.Benzene
24.Fluorobenzene (SS)
25.Trichloroethene
26.1,2-Dichloropropane
27.Bromodichloromethane
28.Dibromomethane
29.2-Chloroethyl vinyl ether
30.cis-1,3-Dichloropropene
31.Toluene
32.trans-1,3-Dichloropropene
33.1,1,2-Trichloroethane
34.2-Bromo-1-chloropropane (IS)
35.1,3-Dichloropropane
Peaks
36.Tetrachloroethene
37.Dibromochloromethane
38.1,2-Dibromoethane
39.Chlorobenzene
40.1,1,1,2-Tetrachloroethane
41.Ethylbenzene
42.m-Xylene
43.p-Xylene
44.1-Chloro-2-fluorobenzene (IS)
45.o-Xylene
46.Styrene
47.Bromoform
48.Isopropylbenzene
49.1,4-Dichlorobutane (SS)
50.1,1,2,2-Tetrachloroethane
51.1,2,3-Trichloropropane
52.n-Propylbenzene
53.Bromobenzene
54.2-Chlorotoluene
55.1,3,5-Trimethylbenzene
56.4-Chlorotoluene
57.tert-Butylbenzene
58.1,2,4-Trimethylbenzene
59.sec-Butylbenzene
60.p-Isopropyltoluene
61.1,3-Dichlorobenzene
62.1,4-Dichlorobenzene
63.Benzyl chloride
64.n-Butylbenzene
65.1,2-Dichlorobenzene
66.1,2-Dibromo-3-Chloropropane
67.4-Bromochlorobenzene (SS)
68.1,2,4-Trichlorobenzene
69.Hexachlorobutadiene
70.Naphthalene
71.1,2,3-Trichlorobenzene
image
GC_EV00005
ColumnRtx-502.2, 75 m, 0.45 mm ID, 2.55 µm (cat.# 10986)
Sample502.2 CAL2000 MegaMix Mixture (cat.# 30431)
502.2 Calibration Mix #1A (cat.# 30439)
502.2 Internal Standard Mix #2 (cat.# 30041)
1-chloro-2-fluorobenzene (cat.# 30040)
4-bromochlorobenzene (cat.# 30230)
2-chloroethyl vinyl ether (cat.# 30265)
1,4-dichlorobutane (cat.# 30227)
methyl tert-butyl ether (MTBE) (cat.# 30402)
and custom mixtures of Freon 113, allyl chloride, chloroprene, and benzyl chloride.
Diluent:RO water
Conc.:20 ppb in 5 mL
Injectionpurge and trap splitless
Purge and Trap
Instrument:Tekmar LSC-3000 Purge and Trap
Trap Type:Vocarb 3000
Purge: 11 min, flow 40 mL/min
Dry Purge: 1 min, flow 40 mL/min
Desorb Preheat Temp.:245 °C
Desorb: 2 min @ 250 °C
Bake: 8 min @ 260 °C
Oven
Oven Temp.:35 °C (hold 6 min) to 115 °C at 11 °C/min (hold 7 min) to 130 °C at 7 °C/min to 220 °C at 9.2 °C/min (hold 4 min)
Carrier GasHe
DetectorμGold Tandem PID/HALL 2000
InstrumentFinnigan 9001
NotesCarrier Gas: constant pressure @ 9 mL/min.

PID: makeup 7 mL/min., purge 7 mL/min. set @ 0.35 mV, base temp. 200 °C
Hall 2000: Rxn gas 25mL/min., Rxn temp. 940 °C, propanol flow 470 μL/min.
AcknowledgementFinnigan 9001 GC, μGold Tandem Photoionization/HALL 2000 Electrolytic Conductivity Detector provided courtesy of Thermo Finnigan GC & GC/MS Division, 2215 Grand Avenue Pkwy, Austin, Texas 78728
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

502.2

,

EPA 502.2

,

Purge and Trap

,

P&T

,

purge

,

trap

,

VOA

,

volatiles

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us