share:

EPA Method 524.2 on Rtx-502.2

Peaks
1.Dichlorodifluoromethane
2.Chloromethane
3.Vinyl chloride
4.Bromomethane
5.Chloroethane
6.Trichlorofluoromethane
7.1,1-Dichloroethene
8.Methylene chloride
9.trans-1,2-Dichloroethene
10.1,1-Dichloroethane
11.2,2-Dichloropropane
12.cis-1,2-Dichloroethene
13.Bromochloromethane
14.Chloroform
15.1,1,1-Trichloroethane
16.1,1-Dichloropropene
17.Carbon tetrachloride
18.Benzene
19.1,2-Dichloroethane
20.Trichloroethene
21.1,2-Dichloropropane
22.Bromodichloromethane
23.Dibromomethane
24.cis-1,3-dichloropropene
25.Toluene
26.trans-1,3-dichloropropene
27.1,1,2-trichloroethane
28.1,3-dichloropropane
29.Tetrachloroethene
30.Dibromochloromethane
Peaks
31.ethylene dibromide
32.Chlorobenzene
33.Ethylbenzene
34.1,1,1,2-tetrachloroethane
35.m-Xylene
36.p-Xylene
37.o-Xylene
38.Styrene
39.Isopropylbenzene
40.Bromoform
41.1,1,2,2-Tetrachloroethane
42.1,2,3-trichloropropane
43.Propylbenzene
44.Bromobenzene
45.1,3,5-trimethylbenzene
46.2-Chlorotoluene
47.4-Chlorotoluene
48.tert-Butylbenzene
49.1,2,4-trimethylbenzene
50.sec-Butylbenzene
51.p-Isopropyltoluene
52.1,3-dichlorobenzene
53.1,4-dichlorobenzene
54.n-Butylbenzene
55.1,2-dichlorobenzene
56.1,2-dibromo-3-chloropropane
57.1,2,4-trichlorobenzene
58.Hexachlorobutadiene
59.Naphthalene
60.1,2,3-trichlorobenzene
image
GC_EV00010
ColumnRtx-502.2, 40 m, 0.18 mm ID, 1.00 µm (cat.# 40915)
SampleDrinking Water VOA MegaMix, 524.2 Rev. 4.1 (cat.# 30052)
Conc.:200 ppb VOA standards in 5mL RO water
Injectionpurge and trap split (split ratio 30:1)
Liner:1mm Split (cat.# 20973)
Inj. Temp.:200 °C
Purge and Trap
Instrument:Tekmar LCS 2000
Trap Type:Vocarb 4000
Purge: 11 min @ ambient, flow 40 mL/min
Desorb Preheat Temp.:250 °C
Desorb: 2 min @ 245 °C, flow 30 mL/min
Interface Connection:injection port
Transfer Line Tubing:Siltek/Sulfinert Tubing(cat.# 22504)
Oven
Oven Temp.:35 °C (hold 4 min) to 150 °C at 6 °C/min to 220 °C at 8 °C/min (hold 8 min)
Carrier GasHe, constant pressure
Linear Velocity:21 cm/sec @ 35 °C
DetectorMS
Mode:Scan
Transfer Line Temp.:250 °C
Tune Type:BFB
Ionization Mode:EI
Scan Range:35-300 amu
InstrumentHP5890 GC & 5971 MSD
NotesOptimized Using Pro ezGC software
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

drinking

,

water

,

purge & trap

,

purge and trap

,

solvents

,

volatiles

,

purge

,

trap

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us