Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides2 (0.32mm ID)
Peaks | |
---|---|
1. | 2,4,5,6-Tetrachloro-m-xylene![]() |
2. | α-BHC (α-HCH)![]() |
3. | γ-BHC (lindane)![]() |
4. | β-BHC (β-HCH)![]() |
5. | δ-BHC (δ-HCH)![]() |
6. | Heptachlor![]() |
7. | Aldrin![]() |
8. | Heptachlor epoxide![]() |
9. | γ-Chlordane![]() |
10. | α-Chlordane![]() |
11. | 4,4'-DDE![]() |
Peaks | |
---|---|
12. | Endosulfan I![]() |
13. | Dieldrin![]() |
14. | Endrin![]() |
15. | 4,4'-DDD![]() |
16. | Endosulfan II![]() |
17. | 4,4'-DDT![]() |
18. | Endrin aldehyde![]() |
19. | Methoxychlor![]() |
20. | Endosulfan sulfate![]() |
21. | Endrin ketone![]() |
22. | Decachlorobiphenyl![]() |

GC_EV00075A
Column | Rtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) |
Sample | Organochlorine Pesticide Mix AB #2 (cat.# 32292) |
Pesticide Surrogate Mix, EPA 8080, 8081 (cat.# 32000) | |
Conc.: | 16-160 pg |
Injection | Direct |
Liner: | Uniliner liner (cat.# 20335) |
Oven | |
Oven Temp.: | 120 °C (hold 1 min) to 300 °C at 9 °C/min (hold 10 min) |
Carrier Gas | He, constant flow |
Flow Rate: | 1.3 mL/min @ 120 °C |
Dead Time: | 1.9 min |
Detector | ECD @ 300 °C |
Notes | ECD with anode purge Head pressure: 8.7psi (constant) See also: Organochlorine Pesticides US EPA Method 8081 on Rtx-CLPesticides (0.32mm ID) GC_EV00075 |