Organochlorine Pesticides by EPA 8081A on Rtx-CLPesticides & Rtx-CLPesticides2
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice

GC_EV00077
Columns | Rtx-CLPesticides 30 m, 0.32 mm ID, 0.50 µm (cat.# 11139) |
and Rtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) | |
Sample | |
Conc.: | 16-160 pg on-column |
Injection | Direct |
Liner: | Drilled Uniliner (hole near top) (cat.# 20335) |
Oven | |
Oven Temp.: | 80 °C (hold 1 min) to 300 °C at 10 °C/min (hold 15 min) |
Carrier Gas | He, constant pressure (8.7 psi, 60.0 kPa) |
Dead Time: | 1.9 min |
Detector | ECD @ 300 °C |
Notes | Flow rate: helium @ 1.3mL/min. @ 120°C |