share:

Organochlorine Pesticides US EPA Method 8081 on Rtx-35

Peaks
1.2,4,5,6-Tetrachloro- m-xylene (surrogate)
2.α-BHC
3.γ-BHC (lindane)
4.β-BHC
5.δ-BHC
6.Heptachlor
7.Aldrin
8.Heptachlor epoxide
9.γ-Chlordane
10.α-Chlordane
11.4,4'-DDE
Peaks
12.Endosulfan I
13.Dieldrin
14.Endrin
15.4,4'-DDD
16.Endosulfan II
17.4,4'-DDT
18.Endrin aldehyde
19.Methoxychlor
20.Endosulfan sulfate
21.Endrin ketone
22.Decachlorobiphenyl (surrogate)
image
GC_EV00079
ColumnRtx-35, 30 m, 0.32 mm ID, 0.50 µm (cat.# 10439)
SampleOrganochlorine Pesticides
InjectionDirect
Oven
Oven Temp.:120 °C (hold 1 min) to 285 °C at 8.5 °C/min (hold 6 min)
Carrier GasHe, constant flow
Flow Rate:2.1 mL/min @ 120 °C
DetectorECD @ 300 °C
NotesECD with anode purge
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us