This chromatogram may contain analytes that are synonyms of your search terms.
Organophosphorus Pesticides by US EPA Method 8140/8141/8141A on Rtx-CLPesticides2

GC_EV00414
Column | Rtx-CLPesticides2, 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324) |
Sample | triphenylphosphate (cat.# 32281) |
tributylphosphate (cat.# 32280) | |
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277) | |
8141 OP Pesticide Calibration Mix B (cat.# 32278) | |
Conc.: | 200 ng/mL |
Injection | |
Inj. Vol.: | 1 µL splitless (hold 1 min) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 3 min) |
Detector | FPD @ 280 °C |
RELATED CHROMATOGRAM LIBRARY SEARCHES
Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.