share:

This chromatogram may contain analytes that are synonyms of your search terms.

Organophosphorus Pesticides US EPA Method 8140/8141/8141A on Rtx-1

Peaks
1.Dichlorvos
2.Hexamethylphosphoramide
3.Mevinphos
4.Trichlorfon
5.TEPP
6.Thionazin (IS)
7.Demeton-o
8.Ethoprop
9.Tributyl phosphate (SS)
10.Dicrotophos
11.Monocrotophos
12.Naled
13.Sulfotepp
14.Phorate
15.Dimethoate
16.Demeton-S
17.Dioxathion
18.Terbufos
19.Fonophos
20.Phosphamidon isomer
21.Diazinon
22.Disulfoton
23.Phosphamidon
24.Dichlorofenthion
25.Methyl parathion
26.Chlorpyrifos methyl
Peaks
27.Ronnel
28.Fenitrothion
29.Malathion
30.Fenthion
31.Aspon
32.Parathion-ethyl
33.Chlorpyrifos
34.Trichloronate
35.Chlorfenvinphos
36.Merphos
37.Crotoxyphos
38.Stirofos
39.Prothiofos
40.merphos oxidation product
41.Fensulfothion
42.Famphur
43.Ethion
44.Sulprofos
45.Carbophenothion
46.Triphenyl phosphate (IS)
47.Phosmet
48.EPN
49.Azinphos-methyl
50.Leptophos
51.Tri-o-cresyl phosphate
52.Azinphos-ethyl
53.Coumaphos
image
GC_EV00415
ColumnRtx-1, 30 m, 0.32 mm ID, 0.50 µm (cat.# 10139)
SampleTriphenylphosphate (cat.# 32281)
Tributylphosphate (cat.# 32280)
8140/8141 OP Pesticide Calibration Mix A (cat.# 32277)
8141 OP Pesticide Calibration Mix B (cat.# 32278)
Diluent:Hexane
Conc.: Total OPP custom standard mix (0.2µg/mL), 1µL US EPA Method 8141A custom standard mixes (200ng/mL)
Injectionsplitless (hold 1.0 min)
Inj. Temp.:250 °C
Oven
Oven Temp.:100 °C to 180 °C at 10 °C/min (hold 2 min) to 300 °C at 18 °C/min (hold 3 min)
Dead Time:0.78 min @ 100 °C
DetectorFPD @ 280 °C
Download PDF

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us