share:

This chromatogram may contain analytes that are synonyms of your search terms.

Volatile Organics by EPA Method 8021A/502.2 on Rtx-502.2

Peaks
1.Dichlorodifluoromethane
2.Chloromethane
3.Vinyl chloride
4.Bromomethane
5.Chloroethane
6.Trichlorofluoromethane
7.1,1-Dichloroethene
8.Allyl chloride
9.Methylene chloride
10.trans-1,2-Dichloroethene
11.Methyl tert-butyl ether
12.1,1-Dichloroethane
13.cis-1,2-Dichloroethene
14.2,2-Dichloropropane
15.Bromochloromethane
16.Chloroform
17.Carbon tetrachloride
18.1,1,1-Trichloroethane
19.1,1-Dichloropropene
20.Benzene
21.1,2-Dichloroethane
22.Fluorobenzene (SS)
23.Trichloroethene
24.Dibromomethane
25.1,2-Dichloropropane
26.Bromodichloromethane
27.1-Bromo-2-chloroethane (SS)
28.cis-1,3-Dichloropropene
29.Toluene
30.Tetrachloroethene
31.trans-1,3-Dichloropropene
32.1,1,2-Trichloroethane
33.Dibromochloromethane
Peaks
34.1,3-Dichloropropane
35.1,2-Dibromoethane
36.Chlorobenzene
37.Ethylbenzene
38.1,1,1,2-Tetrachloroethane
39.m-Xylene
40.p-Xylene
41.1-Chloro-2-fluorobenzene (SS)
42.o-Xylene
43.Styrene
44.Bromoform
45.Isopropylbenzene
46.Bromobenzene
47.n-Propylbenzene
48.1,1,2,2-Tetrachloroethane
49.2-Chlorotoluene
50.1,2,3-Trichloropropane
51.1,3,5-Trimethylbenzene
52.4-Chlorotoluene
53.tert-Butylbenzene
54.1,2,4-Trimethylbenzene
55.sec-Butylbenzene
56.p-Isopropyltoluene
57.1,3-Dichlorobenzene
58.1,4-Dichlorobenzene
59.n-Butylbenzene
60.1,2-Dichlorobenzene
61.1,2-Dibromo-3-Chloropropane
62.2-Bromo-1-chlorobenzene (SS)
63.Hexachlorobutadiene
64.1,2,4-Trichlorobenzene
65.Naphthalene
66.1,2,3-Trichlorobenzene
image
GC_EV00419
ColumnRtx-502.2, 75 m, 0.45 mm ID, 2.55 µm (cat.# 10986)
with Universal “Y” Press-Tight Connector (cat.# 20405)
Sample
Diluent:RO water
Conc.:10 ppb in 5 mL
Injectionpurge and trap splitless
Purge and Trap
Instrument:Tekmar LSC3100 Purge and Trap
Trap Type:Vocarb 3000
Purge: 11 min, flow 40 mL/min
Dry Purge: 1 min, flow 40 mL/min
Desorb Preheat Temp.:245 °C
Desorb: 2 min @ 250 °C
Bake: 8 min @ 260 °C
Interface Connection:direct
Transfer Line Tubing:0.32mm ID Siltek Guard Column (cat.# 10027)
Oven
Oven Temp.:50 °C (hold 2 min) to 70 °C at 2 °C/min to 130 °C at 9 °C/min to 200 °C at 40 °C/min (hold 5 min)
Carrier GasHe
DetectorµGold Tandem PID/Hall
Reaction Gas Flow:7 mL/min
Ionization Potential:0.35 mV
Base Temp.:200 °C
InstrumentFinnigan 9001
NotesCarrier Gas: constant pressure @ ~ 10 mL/min.
Adjust dichlorodifluoromethane to a retention time of 2.28 min. @ 50°C on the Rtx-VGC column.

PID: makeup 7mL/min.
Hall2000: RxnGas 25mL/min., RxnTemp 940°C, Propanol Flow 470 μL/min.

Primary Column: Rtx-VGC, 75m, 0.45mm ID, 2.55μm (cat.# 19409)
AcknowledgementFinnigan 9001 GC, μGold Tandem Photoionization/HALL 2000 Electrolytic Conductivity Detector provided courtesy of Thermo Finnigan GC & GC/MS Division, 2215 Grand Avenue Pkwy, Austin, Texas 78728
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

502.2

,

Purge and Trap

,

P&T

,

PID

,

ELCD

,

purge

,

trap

,

xylene

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us