share:

Polynuclear Aromatic Hydrocarbons US EPA Method 610 on Rtx-CLPesticides2

Peaks
1.Naphthalene
2.Acenaphthylene
3.Acenaphthene
4.Fluorene
5.Phenanthrene
6.Anthracene
7.Fluoranthene
8.Pyrene
Peaks
9.Benz[a]anthracene
10.Chrysene
11.Benzo[b]fluoranthene
12.Benzo[k]fluoranthene
13.Benzo[a]pyrene
14.Indeno[1,2,3-cd]pyrene
15.Dibenzo[a,h]anthracene
16.Benzo[ghi]perylene
image
GC_EV00596
ColumnRtx-CLPesticides2, 30 m, 0.25 mm ID, 0.20 µm (cat.# 11323)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
Diluent:Methylene chloride
Conc.:50 µg/mL
Injection
Inj. Vol.:1.0 µL splitless (hold 2 min)
Liner:Auto SYS Splitless (2mm ID) w/Deact. Wool (cat.# 20829)
Inj. Temp.:330 °C
Oven
Oven Temp.:40 °C (hold 2 min) to 250 °C at 25 °C/min to 265 °C at 5 °C/min to 300 °C at 25 °C/min (hold 4 min)
Carrier GasH2, constant flow
Flow Rate:4 mL/min
DetectorFID @ 350 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PAH

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us