share:

Diesel Fuel on Rtx-5

image
GC_EV00669
ColumnRtx-5, 15 m, 0.25 mm ID, 0.25 µm (cat.# 10220)
Samplediesel fuel #2 composite (cat.# 31093)
Diluent:methylene chloride
Conc.:500 µg/mL
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:4mm Splitless single gooseneck (cat.# 20904)
Inj. Temp.:260 °C
Oven
Oven Temp.:50 °C (hold 1 min) to 330 °C at 20 °C/min
Carrier GasHe, constant pressure
Linear Velocity:76 cm/sec @ 50 °C
DetectorFID @ 330 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

composite

,

fuel #2

,

diesel fuel

,

diesel

,

DRO

,

diesel range organics

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us