share:

Gasoline Range Organics Oxygenates Rtx-VMS

image
GC_EV00679
ColumnRtx-VMS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 19915)
Sample8260B MegaMix calibration mix kit (cat.# 30475)
California oxygenates mix (cat.# 30465)
VOA calibration mix #1 (ketones) (cat.# 30006)
502.2 Calibration mix #1 (gases) (cat.# 30042)
Unleaded gasoline: unweathered (cat.# 30096)
8260A Surrogate mix (cat.# 30240)
8260 Internal standard mix (cat.# 30074)
Non-oxygenated unleaded gasoline (cat.# custom)
Injectionpurge and trap split (split ratio 25:1)
Inj. Temp.:250 °C
Purge and Trap
Instrument:O.I. 4560 Purge and Trap Concentrator
Trap Type:#10 Tenax/silica gel/carbon molecular sieve
Purge: 11 min @ 20 °C, flow 38 mL/min
Desorb: 1.0 min @ 190 °C, flow 32 mL/min
Bake: 10 min @ 210 °C
Oven
Oven Temp.:35 °C (hold 7 min) to 90 °C at 4 °C/min to 220 °C at 45 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.3 mL/min
Dead Time:1.47 min @ 35 °C
DetectorMS
Mode:Scan
Tune Type:PFTBA/BFB
Scan Range:35-260 amu
InstrumentAglient 5971A GC/MS
Notes10 mL sample @ 40°C
AcknowledgementPurge and Trap courtesy of O.I. Analytical.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

oxygenates

,

VOA

,

volatiles

,

gasoline

,

GRO

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us