share:

This chromatogram may contain analytes that are synonyms of your search terms.

Volatiles US EPA Method 624 on Rtx-624

Peaks
1.Chloromethane
2.Vinyl chloride
3.Bromomethane
4.Chloroethane
5.Trichlorofluoromethane
6.1,1-Dichloroethene
7.Methylene Chloride
8.trans-1,2-Dichloroethene
9.1,1-Dichloroethane
10.Bromochloromethane
11.Chloroform
12.1,1,1-Trichloroethane
13.Pentafluorobenzene
14.Carbon Tetrachloride
15.Benzene
16.1,2-Dichloroethane
17.Fluorobenzene
18.Trichloroethene
Peaks
19.1,2-Dichloropropane
20.Bromodichloromethane
21.2-Chloroethyl vinyl ether
22.cis-1,3-Dichloropropene
23.Toluene
24.2-Bromo-1-chloropropane
25.1,1,2-Trichloroethane
26.Tetrachloroethene
27.Dibromochloromethane
28.trans-1,3-Dichloropropene
29.Chlorobenzene
30.Ethylbenzene
31.Bromoform
32.1,4-Dichlorobutane
33.4-Bromofluorobenzene
34.1,1,2,2-Tetrachloroethane
35.1,3-Dichlorobenzene
36.1,4-Dichlorobenzene
37.1,2-Dichlorobenzene
image
GC_EV00701
ColumnRtx-624, 40 m, 0.18 mm ID, 1.00 µm (cat.# 40925)
Sample624 Internal Standard Mix (cat.# 30023)
624 Surrogate Standard Mix (cat.# 30243)
Volatiles MegaMix, EPA Method 624 (cat.# 30497)
Conc.: compounds at 50 ppb (IS @ 40ppb) in 5mL of RO water
Injectionpurge and trap split (split ratio 40:1)
Inj. Temp.:250 °C
Purge and Trap
Instrument:Tekmar LSC-3100 Purge and Trap
Trap Type:Vocarb 3000 (type K)
Purge: 11 min @ ambient, flow 40 mL/min
Dry Purge: 1 min, flow 40 mL/min
Desorb Preheat Temp.:245 °C
Desorb: 2 min @ 250 °C, flow 10 mL/min
Bake: 8 min @ 260 °C
Interface Connection:injection port
Transfer Line Tubing:Silcosteel transfer line, 1mm ID sleeve
Oven
Oven Temp.:50 °C (hold 4 min) to 100 °C at 12 °C/min to 230 °C at 27 °C/min (hold 2 min)
Carrier GasHe, constant flow
Flow Rate:1.1 mL/min
Dead Time:2.06 min @ 50 °C
DetectorAgilent 5971A GC/MS
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Tune Type:PFTBA/BFB
Ionization Mode:EI
Scan Range:35-260 amu
Notes(MCS bypassed using Silcosteel tubing)
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Purgeable

,

Purgeables

,

organics in water

,

halocarbons

,

VOCs

,

drinking water

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us