share:

Chlorinated Disinfection Byproducts, Chlorinated Solvents, and Halogenated Pesticides by EPA 551.1 on Rtx-1301

Peaks
1.Chloroform
2.1,1,1-Trichloroethane
3.Carbon tetrachloride
4.Trichloroacetonitrile
5.Trichloroethylene
6.Bromodichloromethane
7.1,1-Dichloropropanone
8.Chloropicrin
9.Tetrachloroethylene
10.1,1,2-Trichloroethane
11.Dichloroacetonitrile
12.Dibromochloromethane
13.1,2-Dibromoethane
14.1,1,1-Trichloropropanone
15.Bromochloroacetonitrile
16.Bromoform
17.1,2,3-Trichloropropane
Peaks
18.Dibromoacetonitrile
19.1,2-Dibromo-3-Chloropropane
20.Hexachlorocyclopentadiene
21.Trifluralin
22.Hexachlorobenzene
23.Atrazine
24.Simazine
25.γ-BHC (lindane)
26.Heptachlor
27.Alachlor
28.Metolachlor
29.Heptachlor epoxide (B)
30.Bromacil
31.Cyanazine
32.Endrin
33.Endrin aldehyde
34.Endrin ketone
35.Methoxychlor
image
GC_EV00758
ColumnRtx-1301, 30 m, 0.25 mm ID, 1.00 µm (cat.# 16053)
SampleDisinfection Byproducts and Chlorinated Solvents Mix (cat.# 30615)
Method 551.1 Pesticide/Herbicide Mix (cat.# 32438)
Conc.:5-10 µg/mL each analyte
Injection
Inj. Vol.:1.0 µL splitless (hold 0.5 min)
Liner:4mm Split w/Wool (cat.# 20781)
Inj. Temp.:200 °C
Oven
Oven Temp.:35 °C (hold 22 min) to 145 °C at 10 °C/min (hold 2 min) to 225 °C at 20 °C/min (hold 15 min) to 260 °C at 10 °C/min (hold 30 min)
Carrier GasHe, constant pressure
Linear Velocity:30 cm/sec @ 35 °C
DetectorECD @ 290 °C
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

EPA 551.1

,

551.1

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us