share:

This chromatogram may contain analytes that are synonyms of your search terms.

Aroclor PCBs on Rtx-PCB

image
GC_EV00767
ColumnRtx-PCB, 60 m, 0.25 mm ID, 0.25 µm (cat.# 13226)
SampleAroclor 1242 (cat.# 32009)
Aroclor 1254 (cat.# 32011)
Aroclor 1262 (cat.# 32409)
Conc.:333 ppm each
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Splitless taper (4 mm) w/wool (cat.# 22405)
Inj. Temp.:280 °C
Oven
Oven Temp.:100 °C (hold 1 min) to 200 °C at 30 °C/min to 320 °C at 2 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1.1 mL/min
DetectorMS
Mode:Scan
Transfer Line Temp.:280 °C
Ionization Mode:EI
Scan Range:50-550 amu
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Aroclors

,

PCB

,

polychlorinated biphenyls

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us