share:

This chromatogram may contain analytes that are synonyms of your search terms.

Organochlorine Pesticides by EPA Method 8081A on Rtx-440 (dual column w/ Rtx-CLPesticides2)

Peaks
1.Hexachlorocyclopentadiene
2.2,4,5,6-Tetrachloro-m-xylene (SS)
3.cis-Diallate
4.trans-Diallate
5.α-BHC
6.Hexachlorobenzene
7.γ-BHC
8.β-BHC
9.δ-BHC
10.Heptachlor
11.Aldrin
12.Isodrin
13.Heptachlor epoxide
14.2,4'-DDE
15.trans-Chlordane*
16.cis-Chlordane*
Peaks
17.Endosulfan I
18.4,4'-DDE
19.2,4'-DDD
20.Dieldrin
21.Chlorobenzilate
22.Endrin
23.2,4'-DDT
24.4,4'-DDD
25.Endosulfan II
26.Kepone
27.Endrin aldehyde
28.4,4'-DDT
29.Endosulfan sulfate
30.Methoxychlor
31.Endrin ketone
32.Mirex
33.Decachlorobiphenyl (SS)
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV00796
ColumnRtx-440, 30 m, 0.32 mm ID, 0.50 µm (cat.# 12939)
SampleOrganochlorine pesticide mix AB #2, 8-80 µg/mL in ethyl acetate (cat.# 32292)
Chlorobenzilate, 1,000 µg/mL in methanol (cat.# 32211)
Hexachlorobenzene, 1,000 µg/mL in acetone (cat.# 32231)
Hexachlorocyclopentadiene, 1,000 µg/mL in methanol (cat.# 32232)
2,4'-DDD, 1,000 µg/mL in methanol (cat.# 32098)
2,4'-DDE, 1,000 µg/mL in methanol (cat.# 32099)
2,4'-DDT, 1,000 µg/mL in methanol (cat.# 32200)
2,4,5,6-Tetrachloro-m-xylene, 200 µg/mL in acetone (cat.# 32027)
Decachlorobiphenyl (BZ #209), 200 µg/mL in acetone (cat.# 32029)
Diallate (cis & trans), 1,000 µg/mL in hexane (cat.# custom)
Isodrin, 1,000 µg/mL in hexane (cat.# custom)
Kepone, 1,000 µg/mL in hexane (cat.# custom)
Mirex, 1,000 µg/mL in hexane (cat.# custom)
Injection
Inj. Vol.:1.0 µL splitless (hold 0.75 min)
Liner:Siltek single taper (cat.# 20961-214.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:140 °C (hold 1 min) to 240 °C at 30 °C/min (hold 2 min) to 330 °C at 30 °C/min (hold 4 min)
Carrier GasH2, constant pressure
Linear Velocity:51 cm/sec @ 140 °C
DetectorECD @ 330 °C
InstrumentShimadzu 2010 GC
NotesRtx-CLPesticides2 analysis shown in chromatogram GC_EV00797.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

8081A

,

8081

,

organochlorine pesticides

,

OCPs

,

OCP

,

BHC

,

Chlordane

,

Endosulfan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us