share:

This chromatogram may contain analytes that are synonyms of your search terms.

Organochlorine Pesticides on Rtx-CLPesticides Columns (0.18 mm ID column set)

Peaks
1.2,4,5,6-Tetrachloro-m-xylene (SS)
2.α-BHC
3.γ-BHC
4.β-BHC
5.δ-BHC
6.Heptachlor
7.Aldrin
8.Heptachlor epoxide
9.trans-Chlordane*
10.cis-Chlordane*
11.Endosulfan I
Peaks
12.4,4'-DDE
13.Dieldrin
14.Endrin
15.4,4'-DDD
16.Endosulfan II
17.4,4'-DDT
18.Endrin aldehyde
19.Endosulfan sulfate
20.Methoxychlor
21.Endrin ketone
22.Decachlorobiphenyl (SS)
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV00892
ColumnsRtx-CLPesticides 20 m, 0.18 mm ID, 0.18 µm (cat.# 42102)
and Rtx-CLPesticides2 20 m, 0.18 mm ID, 0.14 µm (cat.# 42302)
using IP deactivated guard column 5 m, 0.53 mm ID (cat.# 10045)
with SeCure “Y” connector kit with universal “Y” Press-Tight® connector (cat.# 20276)
SampleOrganochlorine pesticide mix AB #2 (cat.# 32292)
Pesticide surrogate mix, EPA 8080, 8081 (cat.# 32000)
Injection
Inj. Vol.:0.5 µL splitless (hold 0.75 min)
Liner:Splitless taper (2 mm) (cat.# 20796)
Inj. Temp.:250 °C
Oven
Oven Temp.:140 °C (hold 1 min) to 250 °C at 35 °C/min (hold 1 min) to 330 °C at 35 °C/min (hold 3 min)
Carrier GasHe, constant linear velocity
Linear Velocity:20 cm/sec @ 140 °C
DetectorMicro-ECD @ 350 °C
InstrumentAgilent/HP6890 GC
Notescat.# 32292 = 8-80 μg/mL each component in hexane/toluene
cat.# 32000 = 200 μg/mL each component in acetone

This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

Organochlorine pesticides

,

DDT

,

EPA Method 8081

,

Parallel dual column analysis

,

Rtx-CLPesticides

,

BHC

,

Chlordane

,

Endosulfan

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us