Organochlorine Pesticides on Rtx-CLPesticides Columns (0.18 mm ID column set)
Peaks | |
---|---|
1. | 2,4,5,6-Tetrachloro-m-xylene (SS)![]() |
2. | α-BHC![]() |
3. | γ-BHC![]() |
4. | β-BHC![]() |
5. | δ-BHC![]() |
6. | Heptachlor![]() |
7. | Aldrin![]() |
8. | Heptachlor epoxide![]() |
9. | trans-Chlordane*![]() |
10. | cis-Chlordane*![]() |
11. | Endosulfan I![]() |
Peaks | |
---|---|
12. | 4,4'-DDE![]() |
13. | Dieldrin![]() |
14. | Endrin![]() |
15. | 4,4'-DDD![]() |
16. | Endosulfan II![]() |
17. | 4,4'-DDT![]() |
18. | Endrin aldehyde![]() |
19. | Endosulfan sulfate![]() |
20. | Methoxychlor![]() |
21. | Endrin ketone![]() |
22. | Decachlorobiphenyl (SS)![]() |
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice

GC_EV00892
Columns | Rtx-CLPesticides 20 m, 0.18 mm ID, 0.18 µm (cat.# 42102) |
and Rtx-CLPesticides2 20 m, 0.18 mm ID, 0.14 µm (cat.# 42302) | |
using IP deactivated guard column 5 m, 0.53 mm ID (cat.# 10045) | |
with SeCure “Y” connector kit with universal “Y” Press-Tight® connector (cat.# 20276) | |
Sample | Organochlorine pesticide mix AB #2 (cat.# 32292) |
Pesticide surrogate mix, EPA 8080, 8081 (cat.# 32000) | |
Injection | |
Inj. Vol.: | 0.5 µL splitless (hold 0.75 min) |
Liner: | Splitless taper (2 mm) (cat.# 20796) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 140 °C (hold 1 min) to 250 °C at 35 °C/min (hold 1 min) to 330 °C at 35 °C/min (hold 3 min) |
Carrier Gas | He, constant linear velocity |
Linear Velocity: | 20 cm/sec @ 140 °C |
Detector | Micro-ECD @ 350 °C |
Instrument | Agilent/HP6890 GC |
Notes | cat.# 32292 = 8-80 μg/mL each component in hexane/toluene cat.# 32000 = 200 μg/mL each component in acetone This chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector. |