share:

This chromatogram may contain analytes that are synonyms of your search terms.

Pesticides & Herbicides U.S. EPA Method 508.1 Rtx-CLPesticides & Rtx-CLPesticides2

Peaks
1.Hexachlorocyclopentadiene
2.Etridiazole
3.Chlorneb
4.Propachlor
5.Trifluralin
6.Hexachlorobenzene
7.α-BHC
8.Simazine
9.Atrazine
10.Pentachloronitrobenzene (IS)
11.γ-BHC
12.β-BHC
13.δ-BHC
14.Heptachlor
15.Chlorothalonil
16.Metribuzin
17.Alachlor
18.Aldrin
19. 4,4'-Dibromobiphenyl (SS)
Peaks
20.Metolaclor
21.DCPA methyl ester (Chlorthal-dimethyl)
22.Heptachlor epoxide
23.trans-Chlordane*
24.Cyanazine
25.cis-Chlordane*
26.Endosulfan I
27.4,4'-DDE
28.Dieldrin
29.Endrin
30.Chlorobenzilate
31.4,4'-DDD
32.Endosulfan II
33.4,4'-DDT
34.Endrin aldehyde
35.Endosulfan sulfate
36.Methoxychlor
37.cis-Permethrin
38.trans-Permethrin
* For information regarding the nomenclature used for cis-chlordane and trans-chlordane, visit www.restek.com/chlordane-notice
image
GC_EV01022
ColumnsRtx-CLPesticides2 30 m, 0.32 mm ID, 0.25 µm (cat.# 11324)
and Rtx-CLPesticides 30 m, 0.32 mm ID, 0.32 µm (cat.# 11141)
using Rxi guard column 5 m, 0.32 mm ID (cat.# 10039)
with deactivated universal “Y” Press-Tight connector (cat.# 20405-261)
Sample50 ng/mL 508.1 calibration mix #1 (cat.# 32094)
100 ng/mL 508.1 calibration mix #2 (cat.# 32095)
100 ng/mL 508.1 calibration mix #3 (cat.# 32096)
50 ng/mL pentachloronitrobenzene (cat.# 32091)
250 ng/mL 4,4'-dibromobiphenyl (cat.# 32092)
500 ng/mL atrazine (cat.# 32208)
500 ng/mL simazine (cat.# 32236)
Diluent:Ethyl acetate
Injection
Inj. Vol.:2 µL splitless (hold 0.75 min)
Liner:Cyclo double taper (4 mm) (cat.# 20896)
Inj. Temp.:250 °C
Oven
Oven Temp.:80 °C (hold 0.5 min) to 155 °C at 19 °C/min (hold 1 min) to 210 °C at 4 °C/min to 310 °C at 25 °C/min (hold 0.5 min)
Carrier GasHe, constant flow
Linear Velocity:26 cm/sec
DetectorMicro-ECD @ 325 °C
NotesThis chromatogram was obtained using an Agilent micro-ECD. To obtain comparable results, you will need to employ a micro-ECD in addition to confirmational dual columns connected to a 5-meter guard column using a “Y” Press-Tight connector.
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

508.1

,

pesticides

,

herbicides

,

BHC

,

Chlordane

,

Endosulfan

,

52645-53-1

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us