Acephate Rxi-5Sil MS
Peaks | |
---|---|
1. | Acephate![]() |

GC_EV01036
Column | Rxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623) |
using Rxi Guard Column 5 m, 0.25 mm ID (cat.# 10029) | |
Sample | |
Conc.: | 1 ppm each compound |
Injection | |
Inj. Vol.: | 0.5 µL splitless (hold 1 min) |
Liner: | PSS Drilled Uniliner (hole near bottom) (cat.# 22989) |
Inj. Temp.: | 250 °C |
Oven | |
Oven Temp.: | 90 °C (hold 1 min) to 310 °C at 10 °C/min (hold 5 min) |
Carrier Gas | He, constant flow |
Linear Velocity: | 32 cm/sec @ 90 °C |
Detector | MS |
Mode: | Scan |
Transfer Line Temp.: | 300 °C |
Ionization Mode: | EI |
Scan Range: | 50-350 amu |
Instrument | PE Clarus 500 GC & Clarus 500 MS |