share:

This chromatogram may contain analytes that are synonyms of your search terms.

Explosives by US EPA Method 8095 Rxi-5MS with PTV On-Column Liner

image
GC_EV01041
ColumnRxi-5ms, 6 m, 0.53 mm ID, 0.5 µm (cat.# 563153)
using IP Deactivated Guard Column 5 m, 0.53 mm ID (cat.# 10045)
Sample8095 Calibration Mix A (cat.# 31607)
8095 Calibration Mix B (cat.# 31608)
Diluent:acetonitrile
InjectionPTV splitless (hold 0.35 min)
Liner:PTV On-Column Liner (cat.# 24976)
Inlet Temp. Program:55 °C to 285 °C at 10 °C/min (hold 10 min)
PTV Splitless
Purge Flow:15 mL/min
Oven
Oven Temp.:50 °C to 280 °C at 10 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:60 cm/sec @ 300 °C
Detectorµ-ECD @ 300 °C
Make-up Gas Flow Rate:60 mL/min
Make-up Gas Type:N2
NotesAbsolute area reproducibility improves for all compounds, and sensitive compound responses improve dramatically because of the lack of contact with the injection port.

Nitroglycerin: Absolute Area % RSD = 2.6% • Relative Area % RSD = 1.6%
TNT: Absolute Area % RSD = 1.5% • Relative Area % RSD = 1.4%
Download PDF

RELATED CHROMATOGRAM LIBRARY SEARCHES

PTV

,

on-column

,

explosives

,

TNT

,

8095

,

EPA 8095

Printing tips! To print chromatograms full page, change the print settings in the print dialogue window of your PDF reader software: 1) "fit" or "fit to page", 2) "landscape" for wide chromatograms or "portrait" for tall chromatograms.


Restek Domestic Customer Service

Subject

Message

Your Full Name

Your Email

Company Name

Address

Spam Block (Please leave this blank)

all fields required

Thank you

Your message has been sent. We will be in touch shortly.

Message not sent

Sorry, your message could not be sent at this time. Please try again later, or contact Restek or your local Restek representative via phone.

www.restek.com/Contact-Us